5UYM
70S ribosome bound with cognate ternary complex base-paired to A site codon, closed 30S (Structure III)
ELECTRON MICROSCOPY
Sample |
---|
70S ribosome bound with cognate ternary complex base-paired to A site codon, closed 30S (Structure III) |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | GATAN CRYOPLUNGE 3 |
Cryogen Name | ETHANE |
Sample Vitrification Details | 2 uL of complex was applied to each grid. After a 10-second incubation, the grids were blotted for 2 to 4 seconds. |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 153597 |
Reported Resolution (Å) | 3.2 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | ||||
Refinement Space | REAL | ||||
Refinement Protocol | FLEXIBLE FIT | ||||
Refinement Target | correlation coefficient | ||||
Overall B Value | |||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 1 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 500 |
Maximum Defocus (nm) | 5000 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 60976 |
Calibrated Magnification | 60976 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
IMAGE ACQUISITION | SerialEM | 3 |
CTF CORRECTION | CTFFIND | 3 |
MODEL FITTING | UCSF Chimera | 1.7 |
MODEL REFINEMENT | RSRef | 1.1 |
MODEL REFINEMENT | CNS | 1.3 |
MODEL REFINEMENT | PHENIX | 1.11 |
INITIAL EULER ASSIGNMENT | FREALIGN | 9.11 |
FINAL EULER ASSIGNMENT | FREALIGN | 9.11 |
CLASSIFICATION | FREALIGN | 9.11 |
RECONSTRUCTION | FREALIGN | 9.11 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | CTFFIND3 was used to determine CTF values. FREALIGN applied CTF correction. | 800367 | Particles were picked from micrographs using Signature reference-based particle picker. |