6NMI
Cryo-EM structure of the human TFIIH core complex
ELECTRON MICROSCOPY
Sample |
---|
Transcription factor IIH (TFIIH) |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK IV |
Cryogen Name | ETHANE |
Sample Vitrification Details | A thin film of continuous carbon was floated onto Protochips C-flat CF-4/2 holey carbon grids and glow discharged or plasma cleaned before application ... |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 138659 |
Reported Resolution (Å) | 3.7 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (5OF4, 4ERN, 5OQJ, 1G25, 5O85, 2DII, 2JNJ) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | RIGID BODY FIT | ||||
Refinement Target | |||||
Overall B Value | |||||
Fitting Procedure | |||||
Details | Reference structures and homology models were docked and subsequently completely rebuilt according to the density. Parts of the structure were traced ... |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 50 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 500 |
Maximum Defocus (nm) | 700 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | |
Calibrated Magnification | 43478 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details | Residual beam tilt corrected in RELION 3. |
EM Software | ||
---|---|---|
Task | Software Package | Version |
IMAGE ACQUISITION | SerialEM | |
CTF CORRECTION | Gctf | |
CTF CORRECTION | RELION | 3 |
MODEL FITTING | UCSF Chimera | |
MODEL FITTING | Coot | |
MODEL FITTING | O | |
INITIAL EULER ASSIGNMENT | RELION | 3 |
FINAL EULER ASSIGNMENT | RELION | 3 |
CLASSIFICATION | RELION | 3 |
RECONSTRUCTION | RELION | 3 |
MODEL REFINEMENT | PHENIX |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | CTF correction during 3D reconstruction in RELION 3. |