3J8Y
High-resolution structure of ATP analog-bound kinesin on microtubules
ELECTRON MICROSCOPY
Starting Model(s)
Initial Refinement Model(s) | |||
---|---|---|---|
Type | Source | Accession Code | Details |
experimental model | PDB | 4HNA |
Sample |
---|
Microtubule decorated with monomeric human kinesin (K349 construct) having ADP aluminum fluoride complex bound in the nucleotide pocket |
Sample Components |
Human monomeric kinesin-1A construct |
Alpha-tubulin |
Beta-tubulin |
Specimen Preparation | |
---|---|
Sample Aggregation State | FILAMENT |
Vitrification Instrument | HOMEMADE PLUNGER |
Cryogen Name | ETHANE |
Sample Vitrification Details | No glow discharge was applied. After sample application to grid, liquid was mostly 'wicked' away by edgewise application of filter paper. Subsequentl ... |
3D Reconstruction | |
---|---|
Reconstruction Method | HELICAL |
Number of Particles | 49961 |
Reported Resolution (Å) | 5 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | Initial alignment was done using customized SPIDER scripts. Reconstruction and subsequent refinement were done by FREALIGN. |
Refinement Type | |
Symmetry Type | HELICAL |
Axial Symmetry | C1 |
Axial Rise | 8.5215 |
Angular Rotation | 25.77 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (4HNA, 4HNA, 4HNA) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | FLEXIBLE FIT | ||||
Refinement Target | RMSD from the starting structure was monitored for convergence. | ||||
Overall B Value | |||||
Fitting Procedure | |||||
Details | REFINEMENT PROTOCOL--flexible DETAILS--MDFF was performed using explicit solvation, after placing active-site water coordinates identified in high-r ... |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN ULTRASCAN 4000 (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 15 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | 2013-06-02 |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN |
Minimum Defocus (nm) | 1000 |
Maximum Defocus (nm) | 2500 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | GATAN LIQUID NITROGEN |
Nominal Magnification | |
Calibrated Magnification | 23859 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details | 4K x 4K counting mode was used. 24 frames total were collected. |
EM Software | ||
---|---|---|
Task | Software Package | Version |
MODEL FITTING | MDFF | |
RECONSTRUCTION | FREALIGN | |
RECONSTRUCTION | SPIDER |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
done within FREALIGN |