5A63
Cryo-EM structure of the human gamma-secretase complex at 3.4 angstrom resolution.
ELECTRON MICROSCOPY
Sample |
---|
HUMAN GAMMA-SECRETASE |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK IV |
Cryogen Name | ETHANE |
Sample Vitrification Details | VITRIFICATION 1 -- CRYOGEN- ETHANE, HUMIDITY- 100, TEMPERATURE- 85, INSTRUMENT- FEI VITROBOT MARK IV, METHOD- BLOT FOR 4 SECONDS BEFORE PLUNGING, |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 159549 |
Reported Resolution (Å) | 3.4 |
Resolution Method | |
Other Details | SUBMISSION BASED ON EXPERIMENTAL DATA FROM EMDB EMD-3061. (DEPOSITION ID: 13527). |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 38 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | 2014-10-02 |
Temperature (Kelvin) | 85 |
Microscope Model | FEI TECNAI F30 |
Minimum Defocus (nm) | 700 |
Maximum Defocus (nm) | 3200 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | |
Nominal Magnification | 81000 |
Calibrated Magnification | 35714 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
CTF CORRECTION | CTFFIND | 3 |
RECONSTRUCTION | RELION |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
EACH PARTICLE |